Skip to main content
. 2020 Mar 12;20(6):1583. doi: 10.3390/s20061583

Table 10.

Mean respiratory rate, standard deviation comparison for VitalCore with walking step frequency, and respiratory peak accuracy.

Walking Speed Dataset Respiratory Rate (Breaths/Minute) Standard Deviation Step Frequency (Steps/Minute) Peak Detection Accuracy
1 km/h Ground truth 16.32 2.51 64.38 100%
VitalCore 16.53 1.90
2 km/h Ground truth 20.42 3.14 83.04 100%
VitalCore 20.53 3.28
3 km/h Ground truth 17.00 4.23 104.40 88.89%
VitalCore 19.99 7.75
4 km/h Ground truth 21.70 3.02 114.78 100%
VitalCore 21.71 3.13