Skip to main content
. 2020 Feb 27;12(3):203. doi: 10.3390/pharmaceutics12030203

Table 6.

Validation: precision and accuracy data for catechin hydrate (CH) in different biomatrixes. LOQQC, lower limit of quantification for quality control; LQC, lower QC; MQC, medium QC; HQC, high QC.

Biomatrix Quality Controls Samples Theoretical Concentration (ng mL−1) or (ng g−1) Intra-Batch Precision Inter-Batch Precision Recovery (%)
Observed Concentration (ng mL−1) or (ng g−1) ± S.D. Accuracy (%) Precision (%C.V.) Observed Concentration (ng mL−1) or (ng g−1) ± S.D. Accuracy (%) Precision (%C.V.)
Brain Homogenate LOQQC 1.01 0.99 ± 0.03 98.02 3.03 0.98 ± 0.04 97.03 2.90 81.47
LQC 2.90 2.85 ± 0.09 98.28 3.16 2.84 ± 0.11 97.93 2.44 83.48
MQC 410.00 407.61 ± 11.36 99.41 2.79 403.17 ± 12.54 98.33 1.42 80.17
HQC 800.00 789.67 ± 18.64 98.47 3.81 784.69 ± 19.21 98.09 0.72 81.38
Lungs Homogenate LOQQC 1.01 0.98 ± 0.04 97.03 4.08 0.97 ± 0.06 96.04 2.34 80.25
LQC 2.90 2.83 ± 0.11 97.59 3.89 2.81 ± 0.16 96.90 1.04 81.67
MQC 410.00 404.71 ± 13.02 98.71 3.22 402.37 ± 14.17 98.14 0.59 82.49
HQC 800.00 779.25 ± 19.61 97.41 2.52 775.43 ± 20.18 96.93 0.39 81.52
Plasma LOQQC 1.01 0.96 ± 0.03 95.05 3.13 0.94 ± 0.04 93.07 2.32 80.57
LQC 2.90 2.86 ± 0.14 98.62 4.90 2.85 ± 0.15 98.28 1.94 82.61
MQC 410.00 402.33 ± 11.42 98.13 2.84 401.34 ± 12.69 97.89 0.66 83.46
HQC 800.00 778.67 ± 21.64 97.33 2.78 774.18 ± 22.37 96.77 0.39 84.51

Values (mean ± SD) are derived from six replicates.