Analytical imprecision of the Beckman Coulter Access procalcitonin (PCT) immunoassay.

Pools Intra-Assay Inter-Assay Total
n Mean (µg/L) SD (µg/L) CV n Mean (µg/L) SD (µg/L) CV CV
Pool low 20 0.29 0.01 1.8% 10 0.30 0.01 3.6% 4.0%
Pool medium 20 2.83 0.05 1.9% 10 2.85 0.07 2.4% 3.1%
Pool high 20 11.23 0.23 2.1% 10 10.92 0.41 3.7% 4.3%

SD, standard deviation; CV, coefficient of variation.