Table 2.
Classification of electrochemical biosensors employing labels for pathogen detection in terms of: target, working electrode, biorecognition element, electrochemical method, limit of detection, electrochemical probe, and label or secondary processing step. Abbreviations: quartz crystal microbalance (QCM), electrochemical impedance spectroscopy (EIS), cyclic voltommetry (CV), plaque-forming unit (PFU), colony-forming unit (CFU), indium tin oxide (ITO), carbon nanotube (CNT), magnetic bead (MB), nanoparticle (NP), differential pulse voltammetry (DPV), square wave voltammetry (SWV), anodic stripping voltammetry (ASV), hemagglutination units (HAU), and median tissue culture infectious dose (TCID50).
Target Pathogen | Working Electrode | Biorecognition Element | Electrochemical Method & Probe | Limit of Detection | Secondary Binding Step | Reference |
---|---|---|---|---|---|---|
E. coli | ITO electrode | anti-E. coli | EIS; Fe(CN)63-/4- | 6 × 105 cells/mL | antibody/ALP conjugate label for amplification | Yang and Li (2005) |
V. cholerae | carbon/polystyrene electrode | polyclonal anti-V.cholerae | chrono-amperometry | 105 cells/mL | antibody-ALP conjugate label for amplification | Rao et al. (2006) |
E. coli | Au interdigitated microelectrode array | polyclonal anti-E. coli | EIS | 2.67 × 106 cells/mL | antibody-coated MBs for separation | Varshney et al. (2007) |
V. parahaemolytic | carbon electrode | anti-V. parahaemolytic | CV; thionine/hydrogen peroxide | 7.37 × 104 CFU/mL | antibody/HRP conjugate label for transduction | Zhao et al. (2007) |
E. coli | Au interdigitated microelectrode array | polyclonal anti-E. coli | EIS | 7.4 × 104 CFU/mL | antibody-coated MBs for separation and amplification | Varshney and Li (2007) |
E. coli | AuNPs on carbon electrode | monoclonal and polyclonal anti-E. coli | CV; ferrocenedicarboxylic acid/hydrogen peroxide | 6 CFU/mL | polyclonal antibody/HRP conjugate label for amplification | Lin et al. (2008) |
S. aureus | Au electrode | anti-S. aureus | amperometry; tetrathiafulvalene/hydrogen peroxide | 370 cells/mL | antibody/HRP conjugate label for amplification | Escamilla-Gomez et al. (2008) |
S. typhimurium | Au electrode | monoclonal anti-S. typhimurium | chrono-amperometry; tetramethylbenzidine dihydrochloride/hydrogen peroxide | 21 CFU/mL | anti-S. typhimurium polyclonal antibody/HRP conjugate label for amplification | Salam and Tothill (2009) |
S. typhimurium | graphite-epoxy composite electrode | polyclonal anti-S. typhimurium | amperometry | 0.1 CFU/mL | primary antibody-coated MBs for separation, secondary antibody/HRP conjugate label for amplification | Liebana et al. (2009) |
avian influenza virus (AIV) H5N1 | Au interdigitated microelectrode array | monoclonal anti-AIV-H5 | EIS | 0.26 HAU/mL | antibody-coated MBs for separation | Wang et al. (2010) |
Streptococcus pneumoniae | Au electrode | polyclonal anti-S. pneumoniae | amperometry; tetrathiafulvalene/hydrogen peroxide | 1.5 × 104 CFU/mL | antibody-coated MBs for separation and bacteria immobilization, antibody/HRP conjugate label for amplification | Campuzano et al. (2010) |
E. coli | carbon-graphite electrode | monoclonal anti-E. coli | CV | 7 CFU/mL | antibody-coated MBs for separation, antibody/polyaniline label for amplification | Setterington and Alocilja (2011) |
S. aureus | MBs on Au electrode | polyclonal anti-Protein A (S. aureus) | amperometry; tetrathiafulvalene/hydrogen peroxide | 1 CFU/mL | antibody/Protein A/HRP conjugate for amplification | Esteban-Fernandez de Avila et al. (2012) |
avian influenza virus (AIV) H5N1 | Au interdigitated microelectrode array | monoclonal anti-AIV-H5, polyclonal anti-AIV-N1 | EIS | 103 EDI50/mL | anti-AIV-H5 monoclonal antibody- coated MBs for separation, red blood cell label for amplification | Lum et al. (2012) |
E. coli | AuNPs/SiO2 nanocomposite on sulfhydryl chitosan/Fe(C2H5)2/C60 composite on carbon electrode | monoclonal anti-E. coli | CV; ferrocene | 15 CFU/mL | antibody/glucose oxidase/Pt nanochain conjugate label for amplification | Li et al. (2013) |
C. parvum | polypyrrole-coated carbon electrode | polyclonal anti-C. parvum | chrono-potentiometry; o-phenylenediamine/hydrogen peroxide | 500 oocysts/mL | antibody/HRP conjugate label for amplification | Laczka et al. (2013) |
L. monocytogenes | polymeric ion-selective membrane electrode | anti-L. monocytogenes InlA aptamer | potentiometry | 10 CFU/mL | aptamer/protamine label for transduction | Ding et al. (2014) |
avian influenza virus (AIV) H5N1 | Au interdigitated electrode array | anti-AIVH5N1 aptamer | EIS | 0.04 HAU/mL | aptamer-coated MBs for separation, Concanavalin A/glucose oxide-coated AuNP labels for amplification | Fu et al. (2014). |
L. monocytogenes | interdigitated microelectrode array | monoclonal and polyclonal anti-L. monocytogenes | EIS | 300 CFU/mL | monoclonal antibody-coated MBs for separation, polyclonal antibody-coated AuNP label for secondary binding amplification | Chen et al. (2015) |
E. coli | carbon electrode | polyclonal anti-E.coli | chrono-amperometry | 148 CFU/mL | primary antibody-coated MBs for separation, secondary antibody-coated AuNPs for amplification | Hassan et al. (2015) |
avian influenza virus (AIV) H5N1 | AuNPs on ITO microelectrode | polyclonal anti-AIVH5N1 | ASV | 10 pg/mL | antibody-coated MBs for separation and anodic stripping | Zhou et al. (2015) |
E. coli | Au interdigitated microelectrode array | anti-E.coli | EIS; Fe(CN)63-/4- | 100 CFU/mL | wheat germ agglutinin for amplification | Li et al. (2015) |
E. coli | carbon electrode | monoclonal and polyclonal anti-E. coli | DPV | 10 CFU/mL | monoclonal antibody-coated MBs for separation, polyclonal antibody-coated AuNP label for amplification | Wang and Alocilja (2015) |
norovirus | nanostructured Au microelectrode | concanavalin A lectin, polyclonal anti-norovirus | CV, EIS; Fe(CN)63-/4- | 35 copies/mL | antibody-ALP conjugate label for amplification | Hong et al. (2015) |
Legionella pneumophila | carbon electrode | polyclonal anti-L. pneumophila | amperometry; hydroquinone/hydrogen peroxide | 10 CFU/mL | primary antibody- coated MBs for separation, secondary antibody/HRP conjugate label for amplification | Martin et al. (2015) |
S. aureus | carbon electrode | anti-S.aureus aptamer | ASV | 1 CFU/mL | primary aptamer-coated MBs for separation, secondary aptamer-coated AgNP label for anodic stripping | Abbaspour et al. (2015) |
L. monocytogenes | Au interdigitated microelectrode array | monoclonal and polyclonal anti-L. monocytogenes | EIS | 160 CFU/mL | monoclonal antibody-coated MBs for separation, polyclonal antibody-coated AuNP label for amplification | Chen et al. (2016b) |
E. coli | Au interdigitated microelectrode array | polyclonal anti-E. coli | CV, amperometry | 52 CFU/mL | antibody-coated, AuNP/glucose oxidase-modified MBs for separation and amplification | Xu et al. (2016a) |
E. coli | Au interdigitated microelectrode array | anti- E. coli | EIS | 100 CFU/mL | antibody-coated MBs for separation, antibody/glucose oxidase conjugate for amplification | Xu et al. (2016b) |
S. typhimurium | Au interdigitated microelectrode array | monoclonal anti-S. typhimurium | EIS | 100 CFU/mL | antibody-coated MBs for separation, antibody/glucose oxidase conjugate label for amplification | Xu et al. (2016b) |
E. coli | chitosan/CNT composite on carbon electrode | polyclonal anti-E. coli | CV; thionine/hydrogen peroxide | 50 CFU/mL | secondary antibody/HRP conjugate label enzyme-assisted reduction reaction | Gayathri et al. (2016) |
S. typhimurium | carbon electrode | polyclonal and monoclonal anti-S. typhimurium | DPV | 100 cells/mL | polyclonal antibody- coated MBs for separation, monoclonal antibody- coated AuNP label for amplification | Afonso et al. (2016) |
E. coli | Au electrode | anti-E. coli | EIS; Fe(CN)63-/4- | 100 CFU/mL | AuNP label for amplification | Wan et al. (2016) |
L. monocytogenes | Au interdigitated electrode array | polyclonal anti-L. monocytogenes | EIS | 1.6 × 103 CFU/mL | antibody-coated MBs for separation, antibody-coated AuNP label for amplification | Wang et al. (2017) |
E. coli | Au microelectrode | monoclonal anti-E. coli | LSV | 39 CFU/mL | antibody-coated MBs for separation, antibody/AuNP/nucleotide/CdSNP conjugate label for amplification | Li et al. (2017) |
V. cholerae | Au microelectrode | polyclonal anti-V. cholerae | LSV | 32 CFU/mL | antibody-coated MBs for separation, antibody/AuNP/nucleotide/PbSNP conjugate label for amplification | Li et al. (2017) |
avian influenza virus (AIV) H5N1 | Au electrode | anti-AIVH5N1, concanavalin A lectin | CV | 0.367 HAU/mL | Concanavalin A- coated MB labels for amplification | Zhang et al. (2017) |
human influenza A virus H9N2 | carbon electrode | polyclonal anti-influenza A virus M2 protein, fetuin A | chrono-amperometry | 16 HAU | antibody-coated MBs for separation, fetuin A-coated AuNP label for amplification | Sayhi et al. (2018) |
human enterovirus 71 (EV71) | AuNPs on ITO electrode | monoclonal anti-EV71 | CV, EIS, colorimetry; Fe(CN)63-/4- | 10 pg/mL | antibody/HRP-coated MB labels for amplification | Hou et al. (2018) |
E. coli | Ag interdigitated microelectrode array | melittin peptide | EIS | 1 CFU/mL | MLT-coated MBs used for separation and bacteria immobilization | Wilson et al. (2019) |
S. typhimurium | Ag interdigitated electrode array | melittin peptide | EIS | 10 CFU/mL | MLT-coated MBs used for separation and bacteria immobilization | Wilson et al. (2019) |
S. aureus | Ag interdigitated electrode array | melittin peptide | EIS | 110 CFU/mL | MLT-coated MBs used for separation and bacteria immobilization | Wilson et al. (2019) |
Middle East respiratory syndrome corona virus (MERS-CoV) | AuNPs on carbon electrode | MERS-CoV antigen-antibody complex | SWV; Fe(CN)63-/4- | 400 fg/mL | MERS CoV-antibody complex | Layqah and Eissa (2019) |