Skip to main content
. 2020 Apr 1;2020:3012748. doi: 10.1155/2020/3012748

Table 2.

Mean simulated corneal astigmatism and corneal power measurements obtained using IOLMaster, Pentacam, and iTrace.

IOLMaster at 2.3 mm iTrace at 3 mm Pentacam at 3 mm (corneal vertex)
Astigmatism magnitude (D) 1.93 ± 0.88 1.89 ± 0.84 1.88 ± 0.90
Astigmatism cardinal 0.65 ± 1.84 0.74 ± 1.78 0.60 ± 1.79
Astigmatism oblique −0.03 ± 0.86 −0.11 ± 0.78 −0.05 ± 0.89
K flat 44.12 ± 1.69 43.75 ± 1.71 44.06 ± 1.67
K steep 46.05 ± 1.73 45.65 ± 1.72 45.93 ± 1.74
K mean 45.09 ± 1.65 44.71 ± 1.66 45.00 ± 1.65

D = diopters; K = keratometry.