. 2020 Apr 1;2020:3012748. doi: 10.1155/2020/3012748

Table 3.

Comparison of corneal astigmatism values between devices.

Device Correlation between devices (P) Mean difference (D) ± SD 95% CI 95% of LoA spread (D)
Magnitude Cardinal component Oblique component Magnitude Cardinal component Oblique component Magnitude Cardinal component Oblique component Magnitude Cardinal component Oblique component
IOLMaster vs. iTrace 0.853 (<0.0001) 0.965 (<0.0001) 0.920 (<0.0001) 0.03 ± 0.47 −0.09 ± 0.48 0.08 ± 0.34 −0.07, 0.13 −0.19, 0.01 0.01, 0.15 1.84 1.89 1.32
IOLMaster vs. Pentacam 0.870 (<0.0001) 0.957 (<0.0001) 0.839 (<0.0001) 0.05 ± 0.45 0.07 ± 0.53 0.00 ± 0.50 −0.04, 0.15 −0.04, 0.18 −0.10, 0.11 1.77 2.08 1.97
iTrace vs. Pentacam 0.881 (<0.0001) 0.967 (<0.0001) 0.868 (<0.0001) 0.02 ± 0.42 0.16 ± 0.46 −0.08 ± 0.45 −0.07, 0.11 0.07, 0.26 −0.17, 0.01 1.66 1.78 1.76

CI = confidence interval; D = diopters; LoA = limit of agreement; SD = standard deviation.