. 2020 Mar 24;10(3):592. doi: 10.3390/nano10030592

Table 4.

Performance characteristics for the spICP-MS analysis of E171. ESD: equivalent spherical diameter.

ESD Particle Mass Concentration Particle Number Concentration
Repeatability within five series of three repeats (relative standard deviation in percent of average; %) 4.9 18 17
Reproducibility between five series (relative standard deviation in percent of average; %) 8.2 23 13
Size detection limit (nm) 39 - -
Concentration detection limit (ng/L or particles/L) 1 - 50 200
Relative recovery compared (%) 107 96 -
Uncertainty budget for the mean of three repeats analyzed on a single day
Repeatability uncertainty (%) 2.8 10 10
Between-day uncertainty (%) 6.5 23 13
Trueness uncertainty (%) 5.1 11 7
Combined measurement uncertainty (%; k = 1) 8.8 27.5 18
Expanded measurement uncertainty (%; k = 2) 18 55 36

1 Concentration detection limit expressed in the diluted dispersion.