Table 2. Intra- and interobserver reproducibility of RAL measurements by Pentacam.
Group | Intraobserver | Interobserver | |||
---|---|---|---|---|---|
LoA (mm) | ICC | LoA (mm) | ICC | ||
Cataract | −0.25 to 0.20 | 0.986 | −0.50 to 0.48 | 0.941 | |
Non-cataract | −0.26 to 0.26 | 0.995 | −1.03 to 1.16 | 0.914 | |
All | −0.25 to 0.23 | 0.996 | −0.85 to 0.92 | 0.947 |
RAL, radius of anterior lens surface curvature; LoA, the limit of agreement; ICC, the intraclass correlation coefficient.