Face velocity, and surface area and configuration of masks used in the NIOSH NaCl, PFE, BFE, and VFE test methods.

    Face Velocity (cm/sec)
Type Model NIOSH NaCl PFE BFE VFE
N95 FFR A, B, C, D, F 9.3(150 cm2) 3.1(150 cm2) 3.1(150 cm2) 3.1 (150 cm2)
  E 9.3(150 cm2) 3.1(150 cm2) 10.5(45 cm2) 10.5(45 cm2)
Surgical N95 FFR G, H 9.3(150 cm2) 3.1(150 cm2) 3.1(150 cm2) 3.1(150 cm2)
  I 9.3(150 cm2) 3.1(150 cm2) 10.5(45 cm2) 10.5(45 cm2)
Surgical Mask J, K, L 14.2(100 cm2) 5.2(90 cm2) 10.5(45 cm2) 10.5(45 cm2)

Numbers in parentheses show surface area of materials tested.

No shading - Entire mask (150 cm2); Light shading - Mask material (100 cm2)