Skip to main content
. 2020 Mar 30;5(14):7987–8000. doi: 10.1021/acsomega.0c00037

Table 6. Results for the SST Using the Improved Method Reported in Table 5.

system suitability test (SST) obtained value criterion
Prednisolone Reference Solution at the LOQ: 0.05% of the Working Concentration (0.25 μg/mL)
S/N for the prednisolone peak 15 ≥10
Prednisolone Reference Solution at 0.5% of the Working Concentration (2.5 μg/mL)
number of theoretical plates for the prednisolone peak 35,226/m ≥10,000/m
tailing factor for the prednisolone peak 1.01 0.8–1.5
RSD of the prednisolone peak area (n = 3) 0.8% ≤5%
Solution of Prednisolone FSS