. 2010 Jan 7;65(3):235–242. doi: 10.1111/j.1365-2044.2009.06207.x

Figure 5.

Figure 5

 Performance of ventilator iterations at I:E ratios of 1:2 (◆), 1:1 (▪) and 2:1 (). A, B: design (ii): effect of lung compliance on F I o 2. C, D: design (ii): oxygen consumption increases marginally with increased lung compliance (increased ventilator workload). E, F: design (iii): the F I co 2 (a surrogate for NO in this experiment) was maintained within a narrow range (3–3.8%), suggesting that a stable F INO within acceptable safe limits could be delivered by a design of this type if supplied with appropriately diluted NO.