Table 2.
Oxygen consumption and endurance of original pneumatic ventilator for adverse environments (design i) and a conventional commercial pneumatic ventilator under different conditions.
| Ventilator | Tidal volume; ml | Compliance; ml.cmH2 o −1 | Rate; breaths.min−1 | Endurance of standard 340‐l D cylinder | Oxygen consumption; l.min−1 |
|---|---|---|---|---|---|
| Test ventilator | 800 | 60 | 10 | 5 h 0 min | 1.13 |
| Oxylog | 800 | 60 | 10 | 1 h 34 min | 3.62 |
| Test ventilator | 800 | 40 | 10 | 4 h 27 min | 1.27 |
| Oxylog | 800 | 40 | 10 | 1 h 32 min | 3.70 |
| Test ventilator | 650 | 60 | 10 | 5 h 47 min | 0.98 |
| Oxylog | 650 | 60 | 10 | 1 h 50 min | 3.09 |
| Test ventilator | 650 | 40 | 10 | 5 h 50 min | 0.97 |
| Oxylog | 650 | 40 | 10 | 1 h 55 min | 2.96 |