Skip to main content
. 2017 Jul 3;65(2):141–146. doi: 10.1111/lam.12753

Table 2.

Intra‐ and interassay variability in the D1466 real‐time RT‐PCR

Viral copies C t mean Intra‐assay (repeatability) C t mean Interassay (reproducibility)
CV SD CV SD
8·2 × 106 17·70 0·69 0·12 17·76 2·39 0·42
6·1 × 105 21·06 0·10 0·02 21·32 3·55 0·75
3·7 × 105 24·91 0·04 0·01 24·87 1·27 0·31
3·5 × 104 28·15 0·36 0·10 28·03 1·12 0·31
5·1 × 103 30·87 0·16 0·05 30·85 1·24 0·38
1·5 × 103 33·44 0·53 0·18 32·60 2·93 0·95
5·6 × 101 37·03 0·17 0·06 37·59 0·99 0·37

This article is being made freely available through PubMed Central as part of the COVID-19 public health emergency response. It can be used for unrestricted research re-use and analysis in any form or by any means with acknowledgement of the original source, for the duration of the public health emergency.