. 2015 Oct 29;88(1):51–57. doi: 10.1002/jmv.24299

Table I.

Performance Comparison of Individual Targets in Each Test

Real time RT–PCR Multiplex RT–PCR Luminex xTAG
Virus No. true positive specimens No. detected % Sensitivity (95% CI) No. detected % Sensitivity (95% CI) No. detected % Sensitivity (95% CI)
Rhinovirus/Enterovirus 86 57* 66.2 (55.2, 76.1) 75 87.2 (78.2, 93.4) 75 87.2 (78.2, 93.4)
RSV‐B 38 37 97.3 (86.1, 99.5) 37 97.3 (86.1, 99.5) 18 47.3 (30.9, 64.1)
H1N1pdm09 a 35 35 100 (89.9, 100) 35 100 (89.9, 100) 19 54.2 (36.6, 71.1)
PIV‐3 18 17 94.44 (72.6, 99) 18 100 (81.3, 100) 8 44.4 (21.5, 69.2)
RSV‐A 15 15 100 (78, 100) 15 100 (78, 100) 2 13.3 (2, 40.4)
HMPV 7 5 71.4 (29.2, 95.4) 7 100 (58.9, 100) 6 85.7 (42.2, 97.6)
Influenza B 5 5 100 (47.9, 100) 5 100 (47.9, 100) 1 20 (3.3, 71.1)
Corona OC43 6 6 100 (54, 100) 4 66.6 (22.6, 94.6) 6 100 (54, 100)
PIV‐4 9 4 44.4 (13.9, 78.6) 9 100 (66.2, 100) 7 77.7 (40, 96.5)
PIV‐1 2 2 100 (19.2, 100) 2 100 (19.2, 100) 1 50 (8.1 91.83)
Total sample positive 199 175 193 138

*Real time RT‐PCR detect only Rhinovirus.

a

The xTAG® RVP fast kit detects influenza A only and do not have marker to subtype influenza A in H1N1pdm09.

This article is being made freely available through PubMed Central as part of the COVID-19 public health emergency response. It can be used for unrestricted research re-use and analysis in any form or by any means with acknowledgement of the original source, for the duration of the public health emergency.