Skip to main content
. 2011 Feb 15;83(4):568–573. doi: 10.1002/jmv.22031

Table I.

Assay Results and Virus Copy Numbers

No. Immunochromatography RT‐LAMP pwRT‐LAMP pwRT‐LAMP/dried reagent Real‐time RT‐PCR (copies)
1 + + + + 3.73 × 107
2 + + + + 1.12 × 107
3 + + + + 8.41 × 106
4 + + + + 7.24 × 106
5 + + + + 2.02 × 106
6 + + + + 1.70 × 106
7 + + + + 6.77 × 105
8 + + + + 5.33 × 105
9 + + + 5.27 × 105
10 + + + + 2.73 × 105
11 + + + + 1.74 × 105
12 + + + 7.37 × 104
13 + + + 7.04 × 104
14 + + + + 5.04 × 104
15 + + + 4.08 × 104
16 + + + 1.70 × 104
17 + + + 1.39 × 104
18 + + + 8.39 × 103
19 + + + 7.03 × 103
20 + + + 6.51 × 103
21 + + + 3.82 × 103
22 + + + 3.02 × 103
23 + + + 1.01 × 103
24 + + + 8.65 × 102
25 + + + 5.70 × 102
26 + + + 4.89 × 102
27 + + 4.75 × 102
28–56 0
Positivity rate in 27 samples found positive through real‐time RT‐PCR
40.7% (11/27) 100% (27/27) 96.3% (26/27) 100% (27/27)