Skip to main content
. 2006 Feb 2;23(3):323–330. doi: 10.1002/jmri.20504

Table 1.

Comparison of T1 Measurement Techniques in Phantoms

Gd (mM) T1ref (seconds) IR‐ss 4pt (seconds) IR 2pt (seconds)
Meana Meana %diff Meana %diff
0 2.82 2.78 −1.2 1.63 −42.1
0.05 1.86 1.84 −1.2 1.39 −25.6
0.10 1.33 1.35 1.8 1.16 −12.5
0.15 1.06 1.07 0.9 0.90 −15.2
0.20 0.83 0.82 −0.9 0.77 −7.2
0.30 0.62 0.62 0.8 0.59 −3.7
0.40 0.50 0.52 2.8 0.49 −1.8
0.50 0.42 0.44 5.2 0.41 −1.9
0.60 0.36 0.37 3.5 0.36 0.1
0.70 0.31 0.32 5.5 0.31 1.2
0.80 0.28 0.30 5.8 0.28 0.9
0.90 0.25 0.26 3.3 0.25 −0.5
2.00 0.12 0.15 20.7 0.13 1.4
a

SD < 0.01.

IR‐SS 4pt = single‐shot IR b‐SSFP 4‐point (pre‐contrast), IR 2pt = segmented IR b‐SSFP 2‐point ratio method (post‐contrast).