Skip to main content
. 2020 Mar 20;40(3):316–320. [Article in Chinese] doi: 10.12122/j.issn.1673-4254.2020.03.03

1.

灭活前与灭活处理后免疫层析法检测结果

Immunochromatographic results before and after heat inactivation of the samples

Sample Before inactivation n After inactivation Coincidence rate (%) κ P 95CI%
Positive(+) Negative(-)
Serum Positive (+) 11 10 1 95.2 0.905 0.000 0.000-0.000
Negative (-) 10 0 10
Plasma Positive (+) 11 10 1 95.2 0.905 0.000 0.000-0.000
Negative (-) 10 0 10
Whole blood Positive (+) 11 11 0 100.0 1.000 0.000 0.000-0.000
Negative (-) 10 0 10