. 2020 Feb 1;10(2):80. doi: 10.3390/diagnostics10020080

Table 2.

Concurrent validity between the first IMU assessment performed on the first day of measurements and CROM by groups.

Spatial Plane IMU First Assessment on the First Day (Standard Deviation) CROM Assessment (Standard Deviation) Pearson r (p-Value) Student’s t-test (p-Value)
CP group (n = 23)
Flexion-Extension 133.3 (24.6) 129.0 (22.4) 0.969 (<0.001) −3.333 (0.003)
Rotational 153.5 (19.9) 145.3 (14.7) 0.601 (0.003) −2.396 (0.026)
Side-bending 91.3 (25.7) 96.1 (23.3) 0.916 (<0.001) 2.236 (0.036)
Control group (n = 23)
Flexion-Extension 137.0 (24.5) 133.3 (23.9) 0.992 (<0.001) −5.771 (<0.001)
Rotational 147.0 (19.0) 146.1 (16.1) 0.786 (<0.001) −0.371 (0.714)
Side-bending 90.7 (17.5) 94.9 (18.9) 0.949 (<0.001) 3.413 (0.002)

Abbreviations: IMU, Inertial Measurement Unit; CROM, Cervical Range of Motion; CI, confidence interval; CP, cerebral palsy. Evaluation data are expressed in degrees.