. 2009 Feb 2;38(2):230–241. doi: 10.1111/j.1939-165X.2008.00107.x

Table 2.

 Comparison of effusion data between the Sysmex XT‐2000iV and SEAC Hemat 8.

Species Analyte (Units) Sysmex XT2000iV SEAC Hemat 8 P‐Value
(Wilcoxon)
Mean ± SD Median 95th Percentile Range Mean ± SD Median 95th Percentile Range
Dog (n=37) RBC (× 106/μL) 0.95 ± 1.88 0.09 0–5.23 0.91 ± 1.74 0.11 0.01–5.12 .356
HCT (%) 6.79 ± 13.11 0.65 0–42.42 6.32 ± 12.34 0.6 0–35.5 .200
TNCC‐DIFF (× 103/μL) 34.4 ± 86.4 8.6 0.2–239.8 31.8 ± 76.0 8.8 1.0–305.2 .064
TNCC‐BASO (× 103/μL) 35.1 ± 85.4 10.1 1.0–212.4 .004
Cat (n=35) RBC (× 106/μL) 0.10 ± 0.32 0.01 0–0.99 0.1 ± 0.3 0.03 0.01–1.04 .001
HCT (%) 0.53 ± 1.65 0.10 0–5.31 0.49 ± 1.18 0.1 0–3.7 .769
TNCC‐DIFF (× 103/μL) 7.6 ± 13.7 3.0 0.2–40.7 6.8 ± 12.1 2.8 0.4–36.7 .020
TNCC‐BASO (× 103/μL) 7.4 ± 14.2 2.9 0.2–41.4 .082

TNCC, total nucleated cell count.