Skip to main content
. Author manuscript; available in PMC: 2020 Apr 20.
Published in final edited form as: J Magn Reson. 2019 Feb 1;301:73–79. doi: 10.1016/j.jmr.2019.01.015

Table 1:

Summary of coil combination methods with spectroscopy and imaging applications.

Spectroscopy Application Imaging Application
SOS Isos(xi,yj,f)=kKCoil|I(xi,yj,f,ck)|2 Isos(xi,yj,N)=kKCoil|I(xi,yj,N,ck)|2
SVD U × Σ × VH = SVD(I(xi,yj,f,c))
σ1 = max(Σ)
u1 = U·,1
ISVD(xi,yj,f) = σ1u1
N/A
RefPeak S(xi,yj,ck)=I(xi,yj,fRef,ck)ISOS(xi,yj,fRef)
IRefPeak(xi,yj,f)=kKCoilS*(xi,yj,ck)I(xi,yj,f,ck)
S(xi,yj,ck)=I(xi,yj,NRef,ck)IRefsos(xi,yj,NRef)
IRefPeak(xi,yj,N)=kKCoilS*(xi,yj,ck)I(xi,yj,N,ck)
ESPIRiT S(xi,yj,ck)=Ecalib(kspcalib)
IESPIRiT(xi,yj,f)=kKCoilS*(xi,yj,ck)I(xi,yj,f,ck)
S(xi,yj,ck)=Ecalib(kspcalib)
IESPIRiT(xi,yj,N)=kKCoilS*(xi,yj,ck)I(xi,yj,N,ck)

I: original data; xi: ithvoxel in one spatial dimension; yj: jth voxel in the other spatial dimension; f: spectral points; ck and K: kth channel with a total of K channels; N: metabolite in metabolite-specific imaging approach; ∅: arbitrary phase for each voxel; S: sensitivity map; *: conjugate; H: conjugate transpose; fRef: reference peak picked from spectroscopic datasets for coil combination; NRef: reference metabolite image used for imaging datasets coil combination; kspcalib: part of k-space data used for ESPIRiT-derived sensitivity map calculation ISOS, ISVD, IRefPeak, IESPIRiT: combined spectroscopy or imaging datasets with the SOS, SVD, RefPeak, and ESPIRiT methods.