Skip to main content
. 2020 Apr 10;2020:3593805. doi: 10.1155/2020/3593805

Table 7.

Robustness values for the chlorthalidone and the in-house process-related impurities.

Sr. no. Name Control pH Flow rate Wavelength
RRT
pH 5.3 pH 5.7 1.2 mL/min 1.6 mL/min 218 nm 222 nm
1 Chlorthalidone 1.0 1.0 1.0 1.0 1.0 1.0 1.0
2 Stage II process impurity 4.91 4.98 4.89 4.90 4.93 4.92 4.91
3 Stage III process impurity 1.45 1.49 1.42 1.45 1.44 1.46 1.45