Skip to main content
. 2020 Apr 20;20:90. doi: 10.1186/s12871-020-01009-7

Table 4.

Data from intubation in Scenario D: Manual cervical inline stabilization. Data are presented as median (IQR), or as number (percentage)

parameter standard bougie flexible tip bougie catheter p-value
Overall success rate, % 32 (100%) 32 (100%) NS
Duration of 1st intubation attempt, s 34 (30–48) 29 (25–34) 0.001
Number of intubation attempts (%)
 1 27 (84%) 30 (94%)
 2 5 (16%) 2 (6%)
 3 0 (0%) 0 (0%)
Median (IQR) 1 (1–1) 1 (1–1) NS
Number of optimization maneuvers (%)
 0 10 (31%) 12 (37%)
 1 15 (47%) 19 (59%)
 2 7 (22%) 1 (3%)
Median (IQR) 1 (0–1) 1 (0–1) NS
Severity of dental compression (%)
 0 0 (0%) 6 (19%)
 1 12 (37%) 15 (47%)
 2 20 (62%) 11 (34%)
Median (IQR) 2 (1–2) 1 (1–2) 0.015
Ease of use (1–100) 61 (40–72) 53 (38–69) 0.013

NS Not statistically significant