Skip to main content
. 2020 Apr 20;20:90. doi: 10.1186/s12871-020-01009-7

Table 5.

Data from intubation in Scenario E: Cervical collar stabilization. Data are presented as median (IQR), or as number (percentage)

parameter standard bougie flexible tip bougie catheter p-value
Overall success rate, % 32 (100%) 32 (100%) NS
Duration of 1st intubation attempt, s 46 (38–53) 37 (31.5–46) < 0.001
Number of intubation attempts (%)
 1 26 (81%) 30 (94%)
 2 5 (16%) 1 (3%)
 3 1 (3%) 1 (3%)
Median (IQR) 1 (1–1) 1 (1–1) NS
Number of optimization maneuvers (%)
 0 9 (28%) 12 (37%)
 1 15 (47%) 17 (53%)
 2 8 (25%) 3 (9%)
Median (IQR) 1 (0–1.3) 1 (0–1) NS
Severity of dental compression (%)
 0 0 (0%) 4 (12%)
 1 9 (28%) 13 (41%)
 2 23 (72%) 15 (47%)
Median (IQR) 2 (1–2) 1 (1–2) NS
Ease of use (1–100) 72 (53–79) 60 (45–71) 0.014

NS Not statistically significant