Skip to main content
. 2014 May 21;34(2):365–385. doi: 10.1016/j.cll.2014.02.009

Table 3.

Performance of RIDTs with automated readersa

Assay Comparator Flu A Sensitivityb (95% CI) (%) Flu A Specificityb (95% CI) (%) Flu B Sensitivityb (95% CI) (%) Flu B Specificityb (95% CI) (%) Walk-Away
Quidel Sofia Influenza A + B Culture 97 (91–99) 95 (93–96) 90 (83–95) 97 (95–98) Yes
BD Veritor RT-PCR 81.3c (70.0–88.9) 97.4c (94.4–98.8) 85.6c (76.8–91.4) 99.0c (96.5–99.7) No
3M Direct Detection Flu A+ B Culture 80.3 (68.7–89.1) 96.6 (94.8–97.9) 58.3 (27.7–84.8) 98.0 (96.6–99.0) Yes
a

Data derived from product inserts accessed from company Web sites, October 2013.

b

Sensitivity and specificity shown for nasopharyngeal swabs.

c

Positive percent agreement and negative percent agreement shown on product insert.