. 2019 Dec 10;123:104232. doi: 10.1016/j.jcv.2019.104232

Table 3.

Clinical sensitivity and specificity of Panther Fusion® respiratory panel compared to LDT.

Sensitivity
Specificity
N % 95 % CI N % 95 % CI
FLUA 81/86 94.2 [87.1–97.5] 1162/1162 100.0 [99.7–100.0]
FLUB 128/144 88.9 [82.7–93.0] 1101/1104 99.7 [99.2–99.9]
RSV 99/112 88.4 [81.1–93.1] 1136/1136 100.0 [99.7–100.0]
PIV-1 27/33 81.8 [65.6–91.4] 1215/1215 100.0 [99.7–100.0]
PIV-2 23/23 100.0 [85.7–100.0] 1225/1225 100.0 [99.7–100.0]
PIV-3 35/36 97.2 [85.8–99.5] 1211/1212 99.9 [99.5–100.0]
PIV-4 23/32 71.9 [54.6–84.4] 1216/1216 100.0 [99.7–100.0]
AdV 45/57 78.9 [66.7–87.5] 1189/1191 99.8 [99.4–100.0]
RV 160/179 89.4 [84.0–93.1] 1026/1069 96.0 [94.6–97.9]
hMPV 59/66 89.4 [79.7–94.8] 1173/1182 99.2 [98.6–99.6]