Skip to main content
. 2016 Dec 28;242:30–34. doi: 10.1016/j.jviromet.2016.12.016

Table 2.

Overall performance of the multiplex.

Assessment adenovirus astrovirus rotavirus sapovirus
EDL Single 10−3+/− 10−3 10−5 10−6
Multiplex 10−4+/− 10−5+/− 10−5 10−5
R2 Slope R2 Slope R2 Slope R2 Slope
Linearity and efficiency ranges 0.99–1.0 −3.1 → 55 0.99 −3.2 →51 0.99 −3.11 → 27 0.99 −3.1 → 25
Repeatability and reproducibility Mean Ct CV Mean Ct CV Mean Ct CV Mean Ct CV
Intra-assay S3 22.3 0.00479 18.14 0.00529 18.87 0.00903 18.51 0.00656
S2 27.83 0.00829 24.2 0.00788 24.66 0.01525 24.16 0.00612
S1 33.93 0.01847 34.13 0.01037 34.6 0.00887 30.75 0.01106



Inter-assay S3 22.08 0.02397 17.9 0.00582 18.88 0.01497 18.17 0.01221
S2 27.28 0.12473 24.2 0.01727 24.64 0.03103 24.07 0.01248
S1 34.29 0.01735 32.14 0.14854 31.8 0.02164 30.67 0.01364

EDL: endpoint detection limit; R2: reaction efficiency; CV: co-efficient of variation.