Skip to main content
. Author manuscript; available in PMC: 2020 Apr 22.
Published in final edited form as: J Virol Methods. 2018 Jun 30;260:26–33. doi: 10.1016/j.jviromet.2018.06.018

Table 3.

Agreement between laboratories for each kit using Panel 1.

Kit Lab Agreement N (%)
Fleiss’ kappa statistic (95% CI)
All three agreed Two agreed No agreement
InBios 198 (68.5) 86 (29.8) 5 (1.7) 0.62 (0.57, 0.68)
NovaTec 275 (95.2) 11 (3.8) 3 (1.0) 0.91 (0.85, 0.97)
Euroimmun IgM 260 (90.0) 23 (8.0) 6 (2.1) 0.78 (0.72, 0.84)
Euroimmun IgG 270 (93.4) 16 (5.5) 3 (1.0) 0.89 (0.83, 0.95)
Euroimmun IgM & IgG 271 (93.8) 18 (6.2) 0 (0.0) 0.91 (0.85, 0.98)