Table 4.

Diagnostic performance of RTE, TI-RADS, or both in 1525 nodules.

Size interval (cm) AUC (95%CI) Sensitivity (%) Specificity (%) Accuracy (%) PPV (%) NPV (%)
0 < D ≤ 0.5
 TI-RADS 0.916 (0.815–1.000) 89.7 87.5 91.89 91.94 91.67
 RTE 0.832 (0.714–0.950) 74.1 81.2 75.67 93.48 46.43
 TI-RADS plus RTE 0.941 (0.861–1.000) 94.8 87.5 93.24 96.49 82.35
0.5 < D ≤ 1.0
 TI-RADS 0.903 (0.871–0.935) 87.9 85.3 89.20 87.03 94.30
 RTE 0.943 (0.920–0.966) 90.0 93.9 91.48 96.13 84.86
 TI-RADS plus RTE 0.961 (0.941–0.982) 90.0 93.9 91.29 97.03 83.56
1.0 < D ≤ 2.0
 TI-RADS 0.894 (0.869–0.919) 85.0 88.1 83.08 73.11 93.02
 RTE 0.954 (0.937–0.970) 94.1 94.4 94.27 91.81 95.97
 TI-RADS plus RTE 0.972 (0.960–0.985) 94.3 95.6 94.41 92.12 95.98
2.0 < D ≤ 2.5
 TI-RADS 0.942 (0.912–0.973) 86.3 92.9 88.94 94.69 82.11
 RTE 0.919 (0.881–0.956) 79.0 96.4 86.06 97.03 75.70
 TI-RADS plus RTE 0.971 (0.951–0.990) 89.5 95.2 91.83 96.52 86.02

Note. Unless otherwise indicated, the data point is percentage, and the number in parentheses is 95% CI. PPV = positive predictive value; NPV = negative predictive value.