Table 3.
Mean difference (bias) for IOP measurements in pre-surgery with GAT and in post-surgery with GAT, CT1 and CT2.
| IOP difference | GATpost - GATpre | CT1 - GATpre | CT2 - GATpre | |
|---|---|---|---|---|
| All | Mean difference (bias) | −3.569 | 0.324 | 0.912 |
| Wilcoxon test | Z = −8.618; p < 0.001 | Z = −1.318; p = 0.187 | Z = −3.523; p < 0.001 | |
| Regression bias | b = −0.464; p < 0.001 | b = −0.336; p < 0.001 | b = −0.362; p < 0.001 | |
| LASIK | Mean difference (bias) | −3.9452 | −0.192 | 0.425 |
| Wilcoxon test | Z = −7.383; p < 0.001 | Z = −0.823; p = 0.410 | Z = −1.687; p = 0.092 | |
| Regression bias | b = −0.528; p < 0.001 | b = −0.364; p < 0.001 | b = −0.416; p < 0.001 | |
| PRK | Mean difference (bias) | −2.621 | 1.621 | 2.138 |
| Wilcoxon test | Z = −4.373; p < 0.001 | Z = −3.16; p = 0.002 | Z = −3.469; p = 0.001 | |
| Regression bias | b = −0.346; p = 0.010 | b = −0.327; p = 0.035 | b = −0.283; p = 0.108 |
GAT, Goldmann applanation tonometer; CT1-CT2, convex tonometer; IOP intraocular pressure. Z, Wilcoxon test;b, regression coefficient; p, probability value.
Mean difference between measures with GATpre and all the devices in post-surgery were computed and tested for the absence of differences with the Wilcoxon test.
The results of regression bias (related to Bland-Altman) are also shown. Present significant differences are marked in cursive bold.