. 2020 Apr 27;10:7045. doi: 10.1038/s41598-020-63964-y

Table 2.

Interdevice reliability of thromboelastograph parameters between two TEG6S devices.

Parameter Lin’s Concordance Coefficient (95% CI) Reduced Major Axis Intercept Slope
CK R time 0.96 (0.93–0.99) 0.57 0.88
CK K time 0.93 (0.87–0.98) 0.01 1.07
CK alpha 0.87 (0.78–0.96) −14.10 1.20
CK MA 0.99 (0.98–1.00) −1.38 1.03
CK LY30 0.89 (0.82–0.97) 0.07 1.20
HK R time 0.95 (0.92–0.99) 0.33 0.94
HK K time 0.99 (0.98–1.00) 0.04 0.99
HK alpha 0.95 (0.91–0.99) −9.90 1.13
HK MA 1.00 (0.99–1.00) 0.19 1.00
RT R time 0.94 (0.89–0.99) −0.03 0.95
RT K time 1.00 (1.00–1.00) 0.03 0.95
RT alpha angle 0.99 (0.99–1.00) 2.00 0.98
RT MA 1.00 (1.00–1.00) 0.32 1.00
RT LY30 0.95 (0.92–0.98) 0.02 0.79
RT ACT 0.96 (0.93–0.99) −2.96 0.94
FF MA 0.99 (0.98–1.00) −0.55 1.06
FF FLEV 0.99 (0.98–1.00) −13.60 1.07

Abbreviations: CK: standard kaolin; HK: standard kaolin with heparinase; RT: rapid thromboelastograph; R: reaction; K: kinetic; MA: maximum amplitude, LY: amplitude at 30 minutes; ACT: activated clotting time; FF: functional fibrinogen; FLEV: functional fibrinogen level. Reduced major axis slope different to 1 is indicative of the presence of proportional bias; the combination of the slope close to 1 and intercept different from 0 is indicative of the presence of fixed bias.