Table 1.
Different Types Of Waveform Signals Supported By The Monitor.
Wave Type | Sample rate | Sample Period | Sample Size | Array Size | Update Period | Bandwidth Requirement |
---|---|---|---|---|---|---|
ECG | 500 samples/s | 2 ms | 16 bits | 128 samples | 256 ms | 1064 bytes/s |
Compound ECG | 250 samples/s | 4 ms | 16 bits | 3*64 samples | 256 ms | 1640 bytes/s |
Non-ECG waves | 125/62.5 samples/s | 8/16 ms | 16 bits | 32/16 samples | 256 ms | 296/168 byte/s |