Table 1.

Simulated data vs QA-DWI scan results

PVP, % D, *10−3 mm2/s ADC, *10−3 mm2/s Relative difference, %
0 2.25 2.29 1.77
5 2.04 2.06 0.74
10 1.82 1.83 0.35
20 1.51 1.50 0.52
30 1.23 1.24 0.05
40 0.99 1.00 0.11
50 0.82 0.84 1.48
60 0.70 0.70 1.03
70 0.61 0.59 2.17

PVP polyvinylpyrrolidone, D approximated coefficient through mono-exponential equation, ADC apparent diffusion coefficient through DWI