Table 2.
Venous stenosis and occlusion rates following transvenous pacemaker or defibrillator lead insertion.
References | Number of patients and follow-up duration | Definition of stenosis | Baseline stenosis/occlusion rate | New stenosis/occlusion rate |
---|---|---|---|---|
Oginosawa et al. (33) | Total 131 patients. Baseline DSV performed in all. Post implantation DSV performed in 79 patients, mean follow-up 44 ± 6 months | Venous narrowing > 60% of the maximal diameter of the vessel, measured distal to the stenosis site | Stenosis: 6.8% Occlusion: 6.8% |
Stenosis: 13.2% Occlusion: 6.2% |
Abu-El-Haija et al. (32) | Total 150 patients. Baseline DSV performed in all. Post implantation DSV performed in 136 patients at 6 months | Maximum (Dmax) and minimum (Dmin) diameters were measured. Stenosis defined as Dmin/Dmax < 0.4 and Dmin less than 5th centile of all Dmax | Stenosis: 5% Occlusion: 0% |
Stenosis: 10% Occlusion 3.6% |