Skip to main content
. Author manuscript; available in PMC: 2020 May 1.
Published in final edited form as: Nat Protoc. 2019 Dec 20;15(1):161–180. doi: 10.1038/s41596-019-0260-5

Table 1 |.

Troubleshooting table

Step Problem Possible reason Solution
5 No pellet observed after MV isolation A different collection or storage method for plasma (different collection tubes, centrifugation times, among others) was used If plasma is obtained from a third party, make sure the isolation method for plasma is consistent with Steps 1 and 2
An insufficient amount of plasma (<1 mL) was initially used Make sure to use at least 1 mL of plasma
Isolation of plasma after blood draw took too much time, causing loss/degradation of the sample Be sure to isolate the plasma promptly after blood draw
Plasma was frozen and then was not thawed according to the instructions above, causing loss/degradation of the sample Follow the instructions given for thawing of plasma
13 Highly viscous and/or cloudy samples are obtained after lysis Some patients’ plasma is more viscous than others. This can be associated with different diseases A certain level of cloudiness is normal. Sonicate the samples to reduce viscosity before proceeding to the next step and/or increase the EV lysis volume
Samples from different patients were pooled Sonicate the samples to reduce viscosity before proceeding to the next step and/or increase the EV lysis volume
The samples contain a high protein content Sonicate the samples to reduce viscosity before proceeding to the next step and/or increase the EV lysis volume
39 A yellowish sample was obtained after quenching with sodium sulfite The solution was not quenched completely or the sample contains a low amount of glycopeptides Add an extra 2-3 μL of 1 M sodium sulfite and incubate for 10 min with shaking in the dark
49 The StageTips clog If a StageTip flow is slow or stops during glycopeptide desalting, it usually is due to remaining hydrazide beads or any other material in the solution Make sure to avoid taking any remaining material, especially beads, while transferring the eluates from the hydrazide beads. You can also increase the spin time by a couple of minutes and set the centrifugation speed 200g higher
StageTip clogging can also be due to excessive force used to press the material into the tip. Make sure to change packed tip and that the material in the tip is not pressed too tightly.
53 Plasma proteins are highly abundant This usually happens when not all supernatant is removed at the time of EV isolation (Steps 3-11). The MV pellet is usually solid enough to pipette everything out without the pellet being disturbed. The same is true for the exosome pellet, even if this pellet is not visible (most of the time). Make sure to remove all of the supernatant
The yield of glycopeptides and phosphopeptides is low Some common issues that lead to poor recovery are inefficient proteome digestion, use of old buffers, and sample contamination during the enrichment steps If the yield is lower than expected, several parameters should be revised in the data search, including the digestion parameters, and phosphopeptide and glycopeptide enrichment specificity (selectivity). If large numbers of non-phosphorylated or non-glycosylated peptides are observed, it is most likely due to inefficient washing of the beads