Skip to main content
. 2020 Apr 29;12:2979–2992. doi: 10.2147/CMAR.S239927

Table 1.

Summary of Results Before Fine-Tuning in Test Group (111 Cases)

Total n≤3mm 3<n≤4 4<n≤6 n>6 Solid Part-Solid GGO
Detected amount 408 55 115 97 141 261 63 84
True positive 356 46 110 89 111 217 61 78
False positive 52 9 5 8 30 44 2 6
False negative 570 451 82 27 10 445 8 117
Sensitivity 0.3844 0.0926 0.5729 0.7672 0.9174 0.3278 0.8841 0.4
FP per scan 0.6753 0.1169 0.0649 0.1039 0.3896 0.5714 0.026 0.0779
Tested case amount 111
Sensitivity 0.3844

Note: FP per scan: mean false-positive nodules per scan.