. 2020 May 4;20:320. doi: 10.1186/s12879-020-4932-0

Table 1.

Precision of the Deki Reader using laboratory derived specimens

Deki Reader Version RDT brand Result compared to SOC bConcordance n (%)
HIV positive HIV negative Deki & HCW invalid (true invalid)a
True False True False
Intra-instrument reproducibility (5 specimens read 3 times on the same instrument (n = 3 instruments))
V100 (n = 45) c First Response® HIV-1-2-0 25 d 1 16 41 (97.6)
Determine™ HIV-1/2 Ag/Ab Combo 29 16 45 (100)
c,g Uni-Gold HIV 24 9 d 1 33 (97.1)
Overall (n = 121c,g) 78 1 25 1 16 119 (98.3)
Intra-instrument reproducibility (5 specimens read 5 times on the same instrument)
V200 (n = 50) First Response® HIV-1-2-0 25 25 50 (100)
One Step Anti-HIV (1&2) 25 e 1 24 49 (98.0)
HIV 1/2/0 Tri-Line 25 25 50 (100)
Overall (n = 150) 75 1 74 149 (99.3)
Inter-instrument reproducibility (5 specimens read across 3 instruments (3 reads per instrument))
V100 (n = 45) c First Response® HIV-1-2-0 25 d 1 16 41 (97.6)
Determine™ HIV-1/2 Ag/Ab Combo 29 16 45 (100)
c,g Uni-Gold HIV 24 9 e 1 33 (97.1)
Overall (n = 121c,g) 78 1 25 1 16 119 (98.3)
Inter-instrument reproducibility (10 specimens read across 5 instruments)
V200 (n = 50) f First Response® HIV-1-2-0 21 25 46 (100)
One Step Anti-HIV (1&2) 25 24 d 1 49 (98.0)
f HIV 1/2/0 Tri-Line 21 25 46 (100)
Overall (n = 142f) 67 0 74 141 (99.2)

aNumbers indicate concordant invalid results from both Deki Reader and HCW. No false invalid specimens were observed

bConcordance includes invalid specimens, but excludes errors and specimens that were not processed

cSpecimens not uploaded to Fionet™ (First Response® HIV 1–2-0 n = 3; Uni-Gold HIV n = 9)

dPresence of artefact

eHCW transcription error

fSpecimens not captured by HCW (First Response® HIV-1-2-0 n = 4; HIV 1/2/0 Tri-Line n = 4)

g Deki Reader error (n = 2)