. 2020 May 4;20:320. doi: 10.1186/s12879-020-4932-0

Table 2.

HIV-1 and HIV-2 results (RDTs with separate target lines for HIV-1 and HIV-2)

Deki Reader Version RDT brand HIV-1 Result HIV-2 Result
Positive Negative aConcordance
n (%)
Positive Negative a Concordance
n (%)
True False True False True False True False
Intra-instrument reproducibility (5 specimens read 3 times on the same instrument (n = 3 instruments))

bV100

(n = 42)

First Response® HIV-1-2-0 12 0 30 0 42 (100) 13 d 1 28 41 (97.6)
Intra-instrument reproducibility (5 specimens read 5 times on the same instrument)

V200

(n = 50)

First Response® HIV-1-2-0 25 0 25 0 50 (100) 0 d 5 45 0 45 (90.0)
HIV 1/2/0 Tri-Line 25 0 25 0 50 (100) e 10 0 40 0 50 (100)
Inter-instrument reproducibility (5 specimens read across 3 instruments (3 reads per instrument))
bV100 (n = 42) First Response® HIV-1-2-0 12 0 30 0 42 (100) 13 d 1 28 0 41 (97.6)
Inter-instrument reproducibility (10 specimens read across 5 instruments)

cV200

(n = 46)

First Response® HIV-1-2-0 21 0 25 0 46 (100) e 4 d 1 41 0 45 (97.8)
HIV 1/2/0 Tri-Line 21 0 25 0 46 (100) e 6 0 40 0 46 (100)
Laboratory-derived specimen concordance (n = 158)

V200

(n = 158)

First Response® HIV-1-2-0 110 f,g 5 43 0 153 (96.8) 13 g 6 137 f 2 150 (94.9)
HIV 1/2/0 Tri-Line 112 f,g 7 39 0 151 (95.6) 45 g 5 101 f 7 146 (92.4)

aConcordance includes invalid specimens, but excludes errors and specimens that were not processed

bSpecimens not uploaded to Fionet™ (n = 3)

cSpecimens not captured (n = 4)

dPresence of artefact: discordant HIV-2 positive results affecting overall HIV result

ePresence of cross reactivity: concordant HIV-2 positive results (this was linked to two specimens)

fHCW transcription error

gDeki Reader error