Comparison between the CIEDs parameters before and after the examination using BIA.

  Values before BIA assessment Values after BIA assessment P-value
RVI (Ohm) 547.5±97.1 550.3±107.8 0.679
R wave (mV) 11.6±7.0 12.0±6.8 0.225
RVT (V) 0.7±0.5 0.8±0.5 0.276

Data described as mean and standard deviation.

RVI=right ventricle impedance; RVT=right ventricle threshold