Published in final edited form as: J Heart Lung Transplant. 2019 Nov 26;39(5):441–453. doi: 10.1016/j.healun.2019.11.016

Table 7.

Freedom from Adverse Events Categorized by Average Blood Pressure during Axial Flow LVAD Support. Patient average blood pressure was calculated from measures obtained month 3 and onward after LVAD.

Freedom from Incident Event at 3 years, Axial Flow Cohort
Stroke Renal
Failure
Confirmed
Pump
Thrombosis
Infection RV Failure
All Axial Flow 83±0.5% 85±0.4% 95±0.3% 37±0.6% 87±0.5%
Mean Arterial Pressure Sample (n=7046 patients)
Log rank p 0.40 <0.001 0.005 <0.0001 0.27
 MAP ≤75 mmHg (n=1176) 84±1.6% 77±1.4% 94±1.0% 29±1.8% 86±1.5%
 MAP 76-90 mmHg (n=3896) 84±0.8% 84±0.7% 96±0.4% 34±1.0% 86±0.8%
 MAP 91-100 mmHg (n=1496) 84±1.2% 86±1.2% 96±0.7% 36±1.7% 88±1.1%
 MAP >100 mmHg (n=478) 79±2.6% 87±2.0% 95±1.5% 42±2.9% 85±2.6%
Doppler Opening Pressure Sample (n=8619)
Log rank p 0.019 <0.0001 0.016 0.001 0.24
≤80 mmHg (n=2051) 82±1.3% 84±1.2% 95±0.7% 36±1.5% 84±1.3%
81-95 mmHg (n=4783) 85±0.7% 88±0.6% 96±0.4% 40±0.9% 86±0.7%
96-104 mmHg (n=1211) 84±1.4% 87±1.2% 97±0.7% 40±1.8% 86±1.3%
≥105 mmHg (n=572) 79±2.3% 84±2.0% 94±1.4% 43±2.7% 86±2.0%
Systolic Blood Pressure sample (n=8577)
Log rank p 0.001 <0.0001 0.63 <0.0001 0.003
<90 mmHg (n=2150) 84±1.1% 82±1.1% 97±0.6% 31±1.4% 91±0.9%
90-107 mmHg (n=3969) 85±0.7% 86±0.7% 96±0.4% 34±1.0% 89±0.6%
108-119 mmHg (n=1722) 81±1.3% 85±1.1% 96±0.6% 38±1.5% 87±1.2%
≥120 mmHg (n=736) 80±2.1% 85±1.8% 96±1.0% 38±2.4% 85±2.0%

INTERMACS= Interagency Registry of Mechanically Assisted Circulatory Support; LVAD= left ventricular assist device; RV= right ventricular