Table 3.

Studies reporting technical accuracy rates for spine surgery with lumbosacral and thoracic pedicle screw placement using robotic guidance.

Study Robot type No. of pedicles Study design Instrumented spine Technical accuracy, at entry (mm) Angulation error (°)
Togawa, 200730 SpineAssist (Mazor) 32 MIS cadaver setup Lumbar

1.01 ± 0.64 (axial)

0.96 ± 0.62 (sagittal)

Czerny, 201531 iSYS1 (Interventional systems) 20 MIS cadaver setup Thoracic, lumbar

0.35 (axial)

0.7 (sagittal)

Fujishiro, 201532 Renaissance (Mazor) 216 Open surgery cadaver setup Thoracic, lumbosacral

0.64 ± 0.59 (axial)

0.77 ± 0.62 (sagittal)

Lefranc, 201533 ROSA (Medtech) 38 MIS cadaver setup Thoracic, lumbosacral 2.05 ± 1.2 (2D distance)
Van Dijk, 201534 SpineAssist (Mazor) 178 Retrospective MIS patient cohort Lumbar 2.0 ± 1.2 (2D distance)

2.2 ± 1.7 (axial)

2.9 ± 2.4 (sagittal)

Croissant, 201835 MAXIO (Perfint) 24 MIS cadaver setup Thoracic, lumbar

1.2 ± 0.9 (axial)

0.5 ± 0.9 (sagittal)

Jiang, 201836 ExcelsiusGPS (Globus) 8 Retrospective open surgery patient cohort Lumbar 3.2 [0.9-5.4] (2D distance) 2.4 [0.7-3.8] (2D angle)
Current study (Philips) 113 MIS cadaver setup Thoracic and lumbosacral 0.48 ± 0.44 (axial) 0.68 ± 0.58 (sagittal) 0.94 ± 0.83 (axial) 0.87 ± 0.82 (sagittal)

Data are presented as mean ± standard deviation. Abbreviations: MIS = Minimally invasive surgery.