Type of tone |
Pure tone |
Pure tone |
Duration of tone |
2s |
1s |
Response method |
Signal detection, only “yes” responses |
Alternative forced choice, “yes” or “no” |
Possibility of response straight after beginning of tone |
yes |
no |
Duration of response window |
2s |
unlimited, but pop-up window stating < < Choose “no” if you are not sure > > after 3s |
Pause after response |
2 – 3,6 s (random interval) |
0.7s |
Familiarization (right ear) |
1000 Hz, presentation until double confirmation of threshold |
1000 Hz, presentation until double confirmation of threshold |
Familiarization level |
Start value: 40 dB Level decrease for correct response: 10 dB Level increase for wrong response: 5 dB |
Start value: 40 dB Level decrease for correct response: 20 dB Level increase for wrong response: 10 dB |
Criterion for hearing-threshold |
2 confirmations out of 3 presentations of the same ascending threshold at each frequency (shortened ascending method according to DIN EN ISO 8253-1) |
2 confirmations out of 3 presentations of the same ascending threshold at each frequency (shortened ascending method according to DIN EN ISO 8253-1) |
Ascending and descending step size (according to Hughson-Westlake algorithm) |
5 dB HL up, 10 dB HL down |
5 dB HL up, 10 dB HL down |
Default intensity when changing frequency |
10 dB decrease |
20 dB increase |
Default intensity when changing ear |
30 dB |
Hearing threshold from 1000 Hz in the right ear |
Frequency [Hz] order right and left ear |
1000, 1500, 2000, 3000, 4000, 6000, 8000, 1000, 750, 500, 250, 125, (1000) |
1000, 1500, 2000, 3000, 4000, 6000, 8000, 1000, 750, 500, 250, 125 |
Re-Test at 1000 Hz |
Yes |
no |
Procedure if deviation in re-test >10 dB HL |
If difference at 1000 Hz ≤ 5 dB HL, change ear. Otherwise retest all frequencies until difference ≤5 dB HL. |
not applicable |
Class of medical device |
1 |
1 |
Audiometer class |
4 |
4 |