TABLE 4.
Results of comparisons between devices: mean differences (dB HL) of hearing thresholds, p-values and 99.09% CI for the TOST-P in each frequency across all ears.
Frequency [Hz] | Mean diff [dB HL] | t (135) | p | 99.09% CI |
125 | −2.132 | 14.955 | <0.001 | [−3.525, −0.74] |
250 | −2.132 | 21.845 | <0.001 | [−3.086, 1.179] |
500 | −2.721 | 24.331 | <0.001 | [−3.512, −1.929] |
750 | −3.382 | 20.603 | <0.001 | [−4.233, −2.532] |
1000 | −3.235 | 16.263 | <0.001 | [−4.336, −2.134] |
1500 | −4.118 | 14.566 | <0.001 | [−5.187, −3.049] |
2000 | −3.015 | 14.939 | <0.001 | [−4.252, −1.777] |
3000 | −0.478 | 23.984 | <0.001 | [−1.529, 0.573] |
4000 | −0.809 | 21.324 | <0.001 | [−1.95, −0.332] |
6000 | −1.985 | 21.291 | <0.001 | [−3.355, −0.616] |
8000 | −0.919 | 21.028 | <0.001 | [−2.44, 0.602] |