Skip to main content
. 2020 Apr 21;9(4):e17206. doi: 10.2196/17206

Table 1.

Design and assignment of the individual channels of the measuring cartridge.

Cartridge channel Reagents Possible measurements
1
  • Kaolin (intrinsic coagulation)

Clot time
2
  • Kaolin

  • Heparinase (heparin neutralization)

Clot time with heparinase
3
  • Thromboplastin (extrinsic coagulation, tissue factor activation)

  • Polybrene (heparin neutralization)

Clot stiffness
4
  • Thromboplastin (tissue factor activation)

  • Polybrene (heparin neutralization)

  • Abciximab (platelet inhibition)

Fibrinogen contribution
1-2 N/Aa Clot time ratio
3-4 N/A Platelet contribution

aN/A: not applicable.