Skip to main content
. Author manuscript; available in PMC: 2020 May 6.
Published in final edited form as: Mol Imaging Biol. 2019 Apr;21(2):382–390. doi: 10.1007/s11307-018-1231-x

Fig. 4.

Fig. 4.

Results of kinetic parameters of [18F]FLT dynamic PET with shortened acquisitions compared to 30-min acquisition. a Boxplot of Patlak Ki as a function of Td at onset time points (T0) of 1, 2, 3, 4, and 5 min; b correlations of Patlak Ki between shortened and 30-min acquisitions; c boxplot of 2TCM kinetic parameters (Ki, K1, k2, and k3) as a function of Td; and d correlations of 2TCM kinetic parameters between shortened and 30-min dynamic acquisitions. Overall shortened acquisitions with Td ≥ 16 min could enable robust estimate of kinetic parameters.