Table 1.

Scan parameters for the used scan protocols.

Parameter High-pitch “flash” protocol Adaptive sequence “AS” protocol Helical protocol
Scout scans
 Tube voltage 100 kV 100 kV 100 kV
 Tube current 30 mAs 30 mAs 30 mAs

Calcium scoring
 Tube voltage 120 kV 120 kV 120 kV
 Tube current 80 mAs 80 mAs 80 mAs

CCTA
 Heart rate ≤70 bpm 70–90 bpm >90 bpm
 Tube voltage CARE kV CARE kV CARE kV
 Qual. ref. tube voltage 100 kVqual.ref 100 kVqual.ref 100 kVqual.ref
 Qual. ref. tube current 300 mAsqual.ref 300 mAsqual.ref 300 mAsqual.ref
 Slider position 11 11 11
 Delay Test bolus Test bolus Test bolus
Collimation 2  192  0.6 mm 2  160  0.6 mm 2  192  0.6 mm
Rotation time 0.25 s 1.25 s 0.25 s
Pitch 3.2 0.15

Image reconstruction
 Slice thickness 0.6 mm 0.6 mm 0.6 mm
 Increment 0.4 mm 0.4 mm 0.4 mm
 Kernel Bv40 Bv36 Bv40
 Strength 2 3 3

Bpm = beats per minute; CCTA = coronary cardiac computed tomography; Qual. ref. = quality reference.