Subjective surgeon assessment of maneuverability between LithoVue and Uscope for navigation to upper and lower poles when the working channel is empty, with a 1.9-French basket, a 200 μm laser fiber, and a 365 μm laser fiber (upper only).

Subjective scope maneuverability assessment (0–5 rating, 5 = highest) Empty working channel 1.9-French basket 200 μm laser fiber 365 μm laser fiber
Navigation to upper pole
 LithoVue 4.40 (0.51) 4.47 (0.52) 5.00 (0.00) 4.87 (0.35)
 Uscope 3.80 (0.41) 3.47 (0.74) 3.00 (0.65) 2.23 (0.93)
p value <0.01 <0.01 <0.01 <0.01

Navigation to lower pole
 LithoVue 4.33 (0.49) 4.67 (0.49) 5.00 (0.00)
 Uscope 3.40 (0.63) 3.60 (0.51) 3.00 (0.76)
p value <0.01 <0.01 <0.01