Table 2.

Temperature-adjusted (normalized to 21 °C) native T1 and CoV (%) at 3 T summarized by vendor and sequence

Summary table ofCoV ofT1(%) across 4 tubes

Order of tube IDs follows their orientation in the scanned bottle

graphic file with name 12968_2020_613_Figb_HTML.gif

Philips MOLLI 3b(3s)5b 0.33
Siemens ShMOLLI 5b(1b)1b(1b)1b [780C] 0.69
Siemens MOLLI 5b(3s)3b [448B] 0.95
Siemens SASHA 1.29
Philips SASHA 3.76
GE SMART 3.76 sic.
Platform Sequence [prototype, #] ’ CoV ofT1(%), global± SD ofT1according to native tube ID [tube rT1 by slow IR]
M [1260 ms] L [1499 ms] K [1010 ms] J [1872 ms]
Siemens MOLLI MyoMaps product 5s(3s)3s [4] 1.28, 1182 ± 20 1.61, 1367 ± 29 1.02, 955 ± 13 2.42, 1831 ± 62
MOLLI 3b(3s)3b(3s)5b [448, 1] 1.16, 1092 ± 13 1.92, 1252 ± 24 0.56, 889 ± 5 1.58, 1724 ± 27
MOLLI 5b(3s)3b [448B, 1]a 0.36, 1233 ± 4 0.78, 1449 ± 11 0.76, 986 ± 7 0.86, 2012 ± 17
MOLLI 5b(3s)3b [780B, 2] 0.99, 1204 ± 14 1.13, 1410 ± 18 0.74, 970 ± 9 1.50, 1912 ± 28
MOLLI 5s(3s)3s [780B, 1] 1.46, 1222 ± 18 2.36, 1435 ± 34 1.76, 980 ± 17 3.04, 1916 ± 58
ShMOLLI 5b(1b)1b(1b)1b [780B, 2] 1.17, 1155 ± 27 1.62, 1361 ± 35 0.80, 922 ± 17 2.45, 1882 ± 41
ShMOLLI 5b(1b)1b(1b)1b [780C, 1]a 0.78, 1109 ± 9 0.89, 1322 ± 12 0.64, 888 ± 6 1.51, 1892 ± 29
SASHA [2] 2.04, 1284 ± 30 1.15, 1534 ± 27 0.70, 1025 ± 13 1.27, 1949 ± 35
Philips MOLLI 3b(3s)3b(3s)5b [4] 1.64, 1147 ± 52 2.42, 1343 ± 80 1.05, 923 ± 29 3.90, 1785 ± 115
MOLLI 3b(3s)5b [1]a 0.01, 1055 ± 1 0.49, 1217 ± 6 0.05, 862 ± 1 0.78, 1710 ± 13
MOLLI 3s(3s)5s [1] 5.41, 1175 ± 64 3.72, 1392 ± 52 7.13, 937 ± 67 3.56, 1833 ± 65
MOLLI 5b(3s)3b [3] 4.18, 1098 ± 48 5.44, 1286 ± 73 4.37, 868 ± 42 6.33, 1784 ± 106
MOLLI 5s(3s)3s [6] 2.40, 1131 ± 70 2.41, 1320 ± 108 2.60, 906 ± 34 2.71, 1727 ± 131
MOLLI 5b(1b)1b(1b)1bc [2] 2.98, 948 ± 97 4.53, 1082 ± 137 0.96, 794 ± 60 6.29, 1286 ± 282
SASHA [2] 3.11, 1290 ± 46 3.40, 1826 ± 171 2.11, 1054 ± 38 6.41, 2157 ± 267
General Electric SMART [1] 2.85, 1150 ± 33 4.13, 1227 ± 51 2.04, 973 ± 20 6.03, 1419 ± 86

Less favorable CoVs (> 2%, see the “T1 repeatability” section) are in italics. Post-GBCA tubes are not shown here as their data are reported separately in relation to post-GBCA sequences in Supplementary Table 6

sic = text is quoted exactly as it stands in the original, i.e., this is not a typo. Abbreviations as in Table 1

aDenotes the T1 mapping sequence|software combination with lowest overall CoV% for a given vendor where multiples exist

bDenotes the number of different magnets submitting that particular sequence from which the average CoVs were derived

cIn the absence of iterative/data dropping steps in map creation as per ShMOLLI

 = average CoV across the 4 native tubes for a given sequence

’ = where more than one sequence type was submitted, individual CoVs were then averaged to derive x̅’ CoV; while for single sequence submissions x̅’ CoV is from the global mean T1 ± SD for that one sequence