EMG amplitude profiles of recurrent laryngeal nerve and vagus nerve stimulated by the attachable ring stimulator and conventional stimulator.

Patient no. Side V1 (μV) R1 (μV) R2 (μV) V2 (μV)
Ring Conventional Diif Ring Conventional Diif Ring Conventional Diif Ring Conventional Diif
1 Rt 2177 2077 100 2572 2406 166 2361 2460 99 2266 2262 4
Lt 1128 1177 49 1843 1733 110 2178 2530 352 1563 1191 372
2 Lt 1533 1003 530 1103 1773 670 1587 1635 48 996 1073 77
3 Lt 1735 1581 154 2599 2474 125 2658 2487 171 2182 2176 6
4 Lt 1021 1026 5 1375 1392 17 1216 1241 25 864 949 85
5 Rt 641 677 36 730 809 79 770 699 71 743 654 89
6 Rt 1031 977 54 2278 2188 90 1992 1799 193 1214 1111 103
7 Rt 3004 3043 39 2849 2809 40 2636 2583 53 2771 2539 232
Lt 1561 1505 56 1469 1518 49 1741 1584 157 1636 1774 138
8 Lt 762 808 46 541 566 25 834 850 16 561 571 10
9 Rt 621 640 19 716 801 85 1460 1356 104 1219 1157 62
10 Rt 660 653 7 788 786 2 862 902 40 690 689 1
Lt 1026 1058 32 956 981 25 650 603 47 668 634 34
11 Rt 2100 2159 59 2827 2735 92 3639 3633 6 2781 2595 186
12 Lt 607 628 21 1040 1038 2 2403 2040 363 1487 1255 232
13 Lt 849 697 152 1050 727 323 945 818 127 705 546 159
14 Rt 594 525 69 738 578 160 967 953 14 791 757 34
Lt 1146 1195 49 1666 1746 80 1391 1356 35 1177 1084 93
15 Rt 760 792 32 730 693 37 879 784 95 720 768 48
Lt 541 579 38 761 747 14 794 824 30 537 557 20
Mean 1175 1140 77 1432 1425 110 1598 1557 102 1279 1217 99
p value 0.867 0.979 0.876 0.782

Difference between the amplitudes evoked by the attachable ring stimulator and conventional stimulator. p value for the mean amplitudes between the two groups.