Skip to main content
. 2020 Feb 25;92(7):4875–4883. doi: 10.1021/acs.analchem.9b04582

Table 2. Figures of Merit for GC–MS/MS and GC–TOFMS.

    GC–TOFMS
GC–MS/MS m/z 537.2662 m/z 538.2676 m/z 540.2671
LOQa (μmol L–1) 0.01 0.04 0.06 1
sensitivity (ratio peak area analyte/ISTD per μmol L–1) 0.05 0.84 0.37 0.06
linear rangeb (μmol L–1) 0.03–200 0.04–10 0.06–100 1–200
correlation coefficientb 0.9993 (n = 9) 0.9995 (n = 6) 0.9998 (n = 8) 0.9998 (n = 5)
repeatability QC samples (c = 2.5 μmol L–1) 4% (n = 4,16 h) 9% (n = 7, 22 h) 9% (n = 7, 22 h) 5% (n = 7, 22 h)
repeatability samples (extracts) 1% (n = 6, 25 h) 4% (n = 5, 17 h) 4% (n = 5, 17 h) 12% (n = 5, 17 h)
a

Calculated according to Eurachem,24n = 8, c = 0.05 μmol L–1 for GC–MS/MS and c = 0.1 μmol L–1 for TOFMS

b

Corresponding calibration graphs are depicted in Figure S7 for GC–TOFMS and Figure S8 for GC–MS/MS