Table I.

Results of Henry Ford Health System respirator fit testing

FFR model Saccharin solution aerosol fit test performed UVC cycles attempted/fit test cycles passed Passing cumulative UVC dose (3 J/cm2 = 1 cycle)
3M N95 Respirator—1860 NIOSH TC-84A-0006 Baseline, cycles 1-6, 15, 20 & 25 25/20 60 J/cm2
3M N95 Respirator—9210 NIOSH TC-84A-2669 Baseline, cycles 1-2 2/2 6 J/cm2
3M N95 Respirator—8210 NIOSH TC-84A-0007 Baseline, cycles 1-2 2/1 3 J/cm2
Cardinal Health USA N95 R/S Respirator—NIOSH TC-84A-5529 & 5527 (small/regular) Baseline, cycles 1-2 2/1 3 J/cm2
Moldex N95 Respirator #2300N95—NIOSH TC-84A-0328 This N95 respirator passed the baseline fit test on 1 individual. Owing to immediate breakage of straps upon user removal on 2 respirators, testing ceased. 0 N/A
Moldex N95 Respirator 1511 (small)—NIOSH TC-84A-0013 This N95 respirator failed the baseline fit test on 3 individuals. 0 N/A
Moldex N95 Respirator 1512 (medium)—NIOSH TC-84A-0013 Baseline, cycles 1-3 3/2 6 J/cm2
3M N95 Respirator—9010 NIOSH TC-84A-4243 This N95 respirator failed the baseline fit test on 2 individuals. 0 N/A
Cardinal Health USA N95A-S Respirator—NIOSH TC-84A-5463 This N95 respirator failed the baseline fit test on 2 individuals. 0 N/A
GB2626-2206 KN95 Respirator—KN95-01-01 This N95 respirator failed the baseline fit test on 2 individuals. 0 N/A

FFR, Filtering facepiece respirator; N/A, not available; NIOSH, National Institute for Occupational Safety and Health; UVC, ultraviolet C.

3M, St Paul, Minnesota; Cardinal Health, Dublin, Ohio; Moldex-Metric, Culver City, California.

Limited resources prevented testing of additional UVC cycles.